Colorimetric CO2 detector does not turn yellow NRP

For almost a decade now confirmation of intubation is to be done using detection of exhaled CO2. The 7th Edition of NRP has the following to say about confirmation of ETT placement “The primary methods of confirming endotracheal tube placement within the trachea are detecting exhaled CO2 and a rapidly rising heart rate.” They further acknowledge that there are two options for determining the presence of CO2 “There are 2 types of CO2 detectors available. Colorimetric devices change color in the presence of CO2. These are the most commonly used devices in the delivery room. Capnographs are electronic monitors that display the CO2 concentration with each breath.” The NRP program stops short of recommending one versus the other. I don’t have access to the costs of the colorimetric detectors but I would imagine they are MUCH cheaper than the equipment and sensors required to perform capnography using the NM3 monitor as an example. The real question though is if capnography is truly better and might change practice and create a safer resuscitation, is it the way to go?

Fast but not fast enough?

So we have a direct comparison to look at. Hunt KA st al published Detection of exhaled carbon dioxide following intubation during resuscitation at delivery this month. They started from the standpoint of knowing from the manufacturer of the Pedicap that it takes a partial pressure of CO2 of 4 mm Hg to begin seeing a colour change from purple to yellow but only when the CO2 reaches 15 mm Hg do you see a consistent colour change with that device. The capnograph from the NM3 monitor on the other hand is quantitative so is able to accurately display when those two thresholds are reached. This allowed the group to compare how long it took to see the first colour change compared to any detection of CO2 and then at the 4 and 15 mm Hg levels to see which is the quicker method of detection. It is an interesting question as what would happen if you were in a resuscitation and the person intubates and swears that they are in but there is no colour change for 5, 10 or 15 seconds or longer? At what point do you pull the ETT? Compare that with a quantitative method in which there is CO2 present but it is lower than 4. Would you leave the tube in and use more pressure (either PIP/PEEP or both?)? Before looking at the results, it will not shock you that ANY CO2 should be detected faster than two thresholds but does it make a difference to your resuscitation?

The Head to Head Comparison

The study was done retrospectively for 64 infants with a confirmed intubation using the NM3 monitor and capnography.  Notably the centre did not use a colorimetric detector as a comparison group but rather relied on the manufacturers data indicating the 4 and 15 mm Hg thresholds for colour changes.  The mean age of patients intubated was 27 weeks with a range of 23 – 34 weeks.  The results I believe show something quite interesting and informative.

  Median time secs (range)
Earliest CO2 detection 3.7 (0 – 44s)
4 mm Hg 5.3 (0 – 727)
15 mm Hg 8.1 (0 – 727)

I wouldn’t worry too much about a difference of 1.6 seconds to start getting a colour change but it is the range that has me a little worried.  The vast majority of the patients demonstrated a level of 4 or 15 mm Hg within 50 seconds although many were found to take 25-50 seconds.  When compared to a highest level of 44 seconds in the first detection of CO2 group it leads one to scratch their head.  How many times have you been in a resuscitation and with no CO2 change you keep the ETT in past 25 seconds?  Looking closer at the patients, there were 12 patients that took more than 30 seconds to reach a threshold of 4 mm Hg.  All but one of the patients had a heart rate in between 60-85.  Additionally there was an inverse relationship found between gestational age and time to detection.  In other words, the smallest of the babies in the study took the longest to establish the threshold of 4 and 15 mm Hg.

Putting it into context?

What this study tells me is that the most fragile of infants may take the longest time to register a colour change using the colorimetric devices.  It may well be that these infants take longer to open up their pulmonary vasculature and deliver CO2 to the alveoli.  As well these same infants may take longer to open the lung and exhale the CO2.  I suppose I worry that when a resuscitation is not going well and an infant at 25 weeks is bradycardic and being given PPV through an ETT without colour change, are they really not intubated?  In our own centre we use capnometry in these infants (looks for a wave form of CO2) which may be the best option if you are looking to avoid purchasing equipment for quantitative CO2 measurements.  I do worry though that in places where the colorimetric devices are used for all there will be patients who are extubated due to the thought that they in fact have an esophageal intubation when the truth is they just need time to get the CO2 high enough to register a change in colour.

Anyways, this is food for thought and a chance to look at your own practice and see if it is in need of a tweak…

1. Garland J S, Buck R K, Allred E N.et al Hypocarbia before surfactant therapy appears to increase bronchopulmonary dysplasia risk in infants with distress syndrome. Arch Pediatr Adolesc Med 1995149617–622. [PubMed] [Google Scholar]

2. Fujimoto S, Togari H, Yamaguchi N.et al Hypocarbia and cystic periventricular leukomalacia in premature infants. Arch Dis Child Fetal Neonatal Ed 199471F107–F110. [PMC free article] [PubMed] [Google Scholar]

3. Strauss R G. Transfusion therapy in neonates. Am J Dis Child 1991145904–911. [PubMed] [Google Scholar]

4. Hermansen M C, Hermansen M G. Intravascular catheter complications in the neonatal intensive care unit. Clin Perinatol 200532141–156. [PubMed] [Google Scholar]

5. Bhende M S, Thompson A E, Cook D R.et al Validity of a disposable end‐tidal CO2 detector in verifying endotracheal tube placement in infants and children. Ann Emerg Med 199221142–145. [PubMed] [Google Scholar]

6. Noble J J. Carbon‐dioxide monitors:exhaled gas (capnographs, capnometry, end‐tidal CO2 monitors). Pediatr Emerg Care. 1993;9;244–6. [PubMed]

7. Anderson C T, Breen P H. Carbon dioxide kinetics and capnography during critical care. Crit Care 20004207–215. [PMC free article] [PubMed] [Google Scholar]

8. Hagerty J J, Kleinman M E, Zurakowski D.et al Accuracy of a new low‐flow sidestream capnography technology in newborns: a pilot study. J Perinatol 200222219–225. [PubMed] [Google Scholar]

9. Tirosh E, Bilker A, Bader D.et al Capnography in spontaneously breathing preterm and term infants. Clin Physiol 200121150–154. [PubMed] [Google Scholar]

10. Hsieh K S, Lee C L, Lin C C.et al Quantitative analysis of end‐tidal carbon dioxide during mechanical and spontaneous ventilation in infants and young children. Pediatr Pulmonol 200132453–458. [PubMed] [Google Scholar]

11. Repetto J E, Donohue PA‐C P K, Baker S F.et al Use of capnography in the delivery room for assessment of endotracheal tube placement. J Perinatol 200121284–287. [PubMed] [Google Scholar]

12. Roberts W A, Maniscalco W M, Cohen A R.et al The use of capnography for recognition of esophageal intubation in the neonatal intensive care unit. Pediatr Pulmonol 199519262–268. [PubMed] [Google Scholar]

13. Watkins A M, Weindling A M. Monitoring of end tidal CO2 in neonatal intensive care. Arch Dis Child 198762837–839. [PMC free article] [PubMed] [Google Scholar]

14. Wu C H, Chou H C, Hsieh W S.et al Good estimation of arterial carbon dioxide by end‐tidal carbon dioxide monitoring in the neonatal intensive care unit. Pediatr Pulmonol 200335292–295. [PubMed] [Google Scholar]

15. Nangia S, Saili A, Dutta A K. End tidal carbon dioxide monitoring: its reliability in neonates. Indian J Pediatr 199764389–394. [PubMed] [Google Scholar]

16. Rozycki H J, Sysyn G D, Marshall M K.et al Mainstream end‐tidal carbon dioxide monitoring in the neonatal intensive care unit. Pediatrics 1998101648–653. [PubMed] [Google Scholar]

17. Cassady G. Transcutaneous monitoring in the newborn infant. J Pediatr 1983103837–848. [PubMed] [Google Scholar]

18. Boyle R J, Oh W. Erythema following transcutaneous PO2 monitoring. Pediatrics 198065333–334. [PubMed] [Google Scholar]

19. Cabal L A, Hodgman J, Siassi B.et al Factors affecting heated transcutaneous PCO2 and unheated transcutaneous PCO2 in preterm infants. Crit Care Med 19819298 [PubMed] [Google Scholar]

20. Bhat R, Kim W D, Shukla A.et al Simultaneous tissue pH and transcutaneous carbon dioxide monitoring in critically ill neonates. Crit Care Med 19819744–749. [PubMed] [Google Scholar]

21. O'Connor T A, Grueber R. Transcutaneous measurement of carbon dioxide tension during long‐distance transport of neonates receiving mechanical ventilation. J Perinatol 199818189–192. [PubMed] [Google Scholar]

22. Berkenbosch J W, Tobias J D. Transcutaneous carbon dioxide monitoring during high‐frequency oscillatory ventilation in infants and children. Crit Care Med 2002301024–1027. [PubMed] [Google Scholar]

23. Tobias J D, Wilson W R, Jr, Meyer D J. Transcutaneous monitoring of carbon dioxide tension after cardiothoracic surgery in infants and children. Anesth Analg 199988531–534. [PubMed] [Google Scholar]

24. Hand I L, Shepard E K, Krauss A N.et al Discrepancies between transcutaneous and end‐tidal carbon dioxide monitoring in the critically ill neonate with respiratory distress syndrome. Crit Care Med 198917556–559. [PubMed] [Google Scholar]

25. Geven W B, Nagler E, de Boo T.et al Combined transcutaneous oxygen, carbon dioxide tensions and end‐expired CO2 levels in severely ill newborns. Adv Exp Med Biol 1987220115–120. [PubMed] [Google Scholar]

26. Epstein M F, Cohen A R, Feldman H A.et al Estimation of PaCO2 by two noninvasive methods in the critically ill newborn infant. J Pediatr 1985106282–286. [PubMed] [Google Scholar]

27. Aliwalas L L, Noble L, Nesbitt K.et al Agreement of carbon dioxide levels measured by arterial, transcutaneous and end tidal methods in preterm infants <28 weeks gestation. J Perinatol 20052526–29. [PubMed] [Google Scholar]

28. Sivan Y, Eldadah M K, Cheah T E.et al Estimation of arterial carbon dioxide by end‐tidal and transcutaneous PCO2 measurements in ventilated children. Pediatr Pulmonol 199212153–157. [PubMed] [Google Scholar]

29. Tingay D G, Stewart M J, Morley C J. End‐tidal carbon dioxide and transcutaneous carbon dioxide monitoring during neonatal transport. Arch Dis Child Fetal Neonatal Ed 200590F523–F526. [PMC free article] [PubMed] [Google Scholar]

30. McEvedy B A, McLeod M E, Mulera M.et al End‐tidal, transcutaneous, and arterial pCO2 measurements in critically ill neonates: a comparative study. Anesthesiology 198869112–116. [PubMed] [Google Scholar]

31. Bernet‐Buettiker V, Ugarte M J, Frey B.et al Evaluation of a new combined transcutaneous measurement of PCO2/pulse oximetry oxygen saturation ear sensor in newborn patients. Pediatrics 2005115e64–e68. [PubMed] [Google Scholar]

32. Bhende M S. End‐tidal carbon dioxide monitoring in pediatrics: clinical applications. J Postgrad Med 200147215–218. [PubMed] [Google Scholar]

33. Bhende M S, Thompson A E, Howland D F. Validity of a disposable end‐tidal CO2 detector in verifying endotracheal tube position in piglets. Crit Care Med 199119566–568. [PubMed] [Google Scholar]

34. MacLeod B A, Heller M B, Gerard J.et al Verification of endotracheal tube placement with colorimetric end‐tidal CO2 detection. Ann Emerg Med 199120267–270. [PubMed] [Google Scholar]

35. Bhende M S, Thompson A E. Evaluation of an end‐tidal CO2 detector during paediatric cardiopulmonary resuscitation. Pediatrics 199595395–399. [PubMed] [Google Scholar]

36. Varon A J, Morrina J H, Civetta J M. Clinical utility of a colorimetric end‐tidal CO2 detector in cardiopulmonary resuscitation and emergency intubations. J Clin Monit 19917289–293. [PubMed] [Google Scholar]

37. Bhende M S, Allen W D., Jr Evaluation of a Capno‐Flo resuscitator during transport of critically ill children. Pediatr Emerg Care 200218414–416. [PubMed] [Google Scholar]

38. Kamlin C O, O'Donnell C P, Davis P G.et al Colorimetric end‐tidal carbon dioxide detectors in the delivery room: strengths and limitations. A case report. J Pediatr 2005147547–548. [PubMed] [Google Scholar]

39. Aziz H F, Martin J B, Moore J J. The pediatric disposable end‐tidal carbon dioxide detector role in endotracheal intubation in newborns. J Perinatol 199919110–113. [PubMed] [Google Scholar]

40. Niermeyer S, Carlo W, Boyle D.et al What is on the horizon for neonatal resuscitation? Neonatal Resuscitation Program Steering Committee. NeoReviews 20012e51–e57. [Google Scholar]

41. Bhende M S, Thompson A E. Evaluation of an end‐tidal CO2 detector during pediatric cardiopulmonary resuscitation. Pediatrics 199595395–399. [PubMed] [Google Scholar]


Page 2

Colorimetric CO2 detector does not turn yellow NRP

Figure 1 Colorimetric end tidal CO2 (ETco2) detectors: possible causes of apparent false positive and negative results in neonates. VA ECMO, Venoarterial extracorporeal membrane oxygenation.

  • Colorimetric CO2 detector does not turn yellow NRP
  • Colorimetric CO2 detector does not turn yellow NRP

Click on the image to see a larger version.